Salesforce

i7100 - How to Recalibrate Sensor Values

« Go Back

Content

 
Article Number000007161
Titlei7100 - How to Recalibrate Sensor Values
Body
i7100 Sensor Calibration Utility

The i7100 may require sensor calibration after updating to firmware version 5.14 or higher. This should not be required if updating from a firmware version later than 5.14. This sensor calibration can also be performed if the printer has difficulty maintaining calibration despite having configured proper calibration settings: i7100 Sensor Position for Gapped, Sleeve, Notched, and Multiple Column Gapped Material.

This process re-exposes the sensors to the light and dark conditions of a passing label. These instructions must be followed exactly or sensor values will be out of range. For Brady labels, sensor values should fall within this range: Lowest (1-10) - Mid (10-20) / Mid (10-20) - Highest (120-160).

 
i7100 Calibration

Note: The sensor calibration utility can be found under Settings > Setup > Labels > Calibrate Label Sensor. This setting may also be in Setup > Extras depending on your firmware version.
 
  1. On the printer screen, select the Gear icon.
Menu_Settings_Resized.png
 
  1. Select Extras and then choose Calibrate sensor.
    Note: The Calibrate Sensor option may be under Setup > Labels > Calibrate Label depending on your firmware version.
    Extras_Resized.png
 
  1. Remove the labels from under the sensor (orange light), then close the print head and press Continue.
image.png
 
 
i7100 with labels removed and print head open.
image.png
 
  1. Printer will say Insert Liner, remove a couple of labels from the liner and place the blank liner under the sensor (orange light).
image.png
 
image.png
 
  1. Press ContinueThe sensor is now calibrated and you can reload the labels and press Feed a couple of times until one label feeds out each time.

Related information 
URL NamePressing-FEED-causes-printer-to-just-spit-out-labels-after-upgrading-firmware-from-5-08-to-5-14-version

Powered by